The trouble analysis and troubleremoval of wavelength system in MPF-4 Fluorescence Spectrophotometer were introduced in this paper.
本文介绍了MPF—4荧光光谱仪波长系统的故障分析及排除方法。
2
Open a box to cover an inspection, discover almost all thrum all not impaction, one loosens actually, classics the troubleremoval after reinstalling.
打开盒盖检查,发现几乎所有的线头均未压紧,有一根竟然松脱出来,经重新安装后故障排除。
3
Troubleremoval for insufficient oil supply or absence of oil supply caused by poor separation of oil and gas and oil leak from gas outlet is described.